Certified by TUV according to Safety Cabinets European standard EN 12469, Class II Telstar BIO II A provides a high level of protection to operator, environment and products. It is suitable for the manipulation of pathogenes of categories 1, 2 and 3 as stated in the standard but is also the ideal equipment to protect fragile samples from direct or cross contamination.
Taking advantage of the long experience of Telstar in design of equipment for pharmaceutical production, BIO II A offers safety features, options and accessories beyond the standard requirements, to fulfill virtually all needs. Each cabinet is factory-certified by TÜV-GS registered laboratory to meet the specified performance requirements.
The BIO II A Class II cabinet recirculates 70% of the total flow through a HEPA/ULPA filter H14 providing a class 10 laminar flow in the chamber ( 99.999% efficiency for particles > 0.3 μm) and expels the other 30% at top of the unit through a second H14 HEPA/ULPA filter to protect the environment.
Features:
• High downflow air speed (0.4 m/s) and high air inlet speed at front (0.5 to 0.7 m/s) provide both reinforced protection for operator and samples, beyond the regular cabinet design, thanks to a stronger air barrier and to a reduced risk of cross contamination.
• Chamber flow and exhaust flow are controlled by interactive but independent plenums, blowers and sensors. It provides an automatic compensation of any cause of flow variation and permits the connection of accessories at exhaust duct without disturbance of the flows.
• One piece stainless steel chamber, round corners for easy cleaning and decontamination.
• Partitioned stainless steel work bench, easy removable for cleaning and autoclavable.
• Stainless steel armrest resisting to frequent cleaning/disinfection.
• High lighting level of 1100 lux.
• Tempered window glass of safety grade, counterbalanced with gas springs to stand in open position for introduction of large items and easy access for cleaning.
• Cabinet frame of epoxy/polyester painted steel.
• UV light as standard feature with safety interlock with front window.
• All maintenance and filter replacement is easily done from the front of the unit.
• Vacuum tap, gas tap with solenoid valve, electrical socket as standard.
• Possibility to connect an output relay for external fan activation (ducted cabinets).
• Special model with lead protection for radioisotopes preparation.
Specification
Certified by TUV according to Safety Cabinets European standard EN 12469, Class II Telstar BIO II A provides a high level of protection to operator, environment and products. It is suitable for the manipulation of pathogenes of categories 1, 2 and 3 as stated in the standard but is also the ideal equipment to protect fragile samples from direct or cross contamination.
Taking advantage of the long experience of Telstar in design of equipment for pharmaceutical production, BIO II A offers safety features, options and accessories beyond the standard requirements, to fulfill virtually all needs. Each cabinet is factory-certified by TÜV-GS registered laboratory to meet the specified performance requirements.
The BIO II A Class II cabinet recirculates 70% of the total flow through a HEPA/ULPA filter H14 providing a class 10 laminar flow in the chamber ( 99.999% efficiency for particles > 0.3 μm) and expels the other 30% at top of the unit through a second H14 HEPA/ULPA filter to protect the environment.
Features:
• High downflow air speed (0.4 m/s) and high air inlet speed at front (0.5 to 0.7 m/s) provide both reinforced protection for operator and samples, beyond the regular cabinet design, thanks to a stronger air barrier and to a reduced risk of cross contamination.
• Chamber flow and exhaust flow are controlled by interactive but independent plenums, blowers and sensors. It provides an automatic compensation of any cause of flow variation and permits the connection of accessories at exhaust duct without disturbance of the flows.
• One piece stainless steel chamber, round corners for easy cleaning and decontamination.
• Partitioned stainless steel work bench, easy removable for cleaning and autoclavable.
• Stainless steel armrest resisting to frequent cleaning/disinfection.
• High lighting level of 1100 lux.
• Tempered window glass of safety grade, counterbalanced with gas springs to stand in open position for introduction of large items and easy access for cleaning.
• Cabinet frame of epoxy/polyester painted steel.
• UV light as standard feature with safety interlock with front window.
• All maintenance and filter replacement is easily done from the front of the unit.
• Vacuum tap, gas tap with solenoid valve, electrical socket as standard.
• Possibility to connect an output relay for external fan activation (ducted cabinets).
• Special model with lead protection for radioisotopes preparation.
Specification
Model | BIO II A/P | BIO II A | BIO II A/M | BIO II A/G |
Filter efficiency | Efficiency >99.999% (DOP) / >99.995% (MPPS EN 1822) | |||
Noise | ≤59dB(A) | |||
Illumination (lux) | 760 | 930 | 1100 | 1100 |
Air volume (m3/h) | 605 | 995 | 1100 | 1500 |
Air velocity (m/s) | 0.40 | 0.40 | 0.35 | 0.40 |
External dim. (mm) | 852x760x1410 | 1310x760x1410 | 1615x760x1410 | 1920x760x1410 |
Internal dim. (mm) | 732x580x700 | 1190x580x700 | 1495x580x700 | 1800x580x700 |
Power (kw) | 1.2 | 1.8 | 1.8 | 2.9 |
Weight (kg) | 160 | 200 | 235 | 270 |